Peculiarity when measuring UVC LEDs

Ultraviolet optical radiation (UV radiation) is used in a variety of tasks. It can be used to modify the physical properties of materials - radiation curing - or to treat skin diseases such as psoriasis. Water, air and solid surfaces are disinfected with UV radiation. A wide variety of gases and biomolecules can be detected with UV radiation. Among artificial radiation sources of UV radiation, UV LEDs are becoming a serious alternative to the hitherto dominant gas discharge lamps. Although the technology of UV LEDs is still young, it is making fast progress particularly in the UVB and UVC spectral ranges. The classic instrument for measuring the total radiant power of UV LEDs is a spectroradiometer with an integrating sphere.

A peculiarity of quasi monochromatic LED radiation must be considered when designing integrating sphere spectroradiometers for UV LEDs. The monochromatic UV radiation of the LED causes fluorescence of the integrating sphere coating which leads to significant measurement errors that cannot be compensated by conventional correction methods. Research at the Gigahertz-Optik laboratory has confirmed that barium sulfate coatings and synthetic coatings for integrating spheres both fluoresce when exposed to UVC LEDs < 250 nm. This fluorescence is caused by organic molecules which deposit over time on the integrating sphere’s coating. When stimulated with UV-LED radiation <= 230 nm, the radiant flux of the fluorescence can even reach the magnitude of the radiant power of UV-LEDs themselves. The fluorescence level can be reduced by irradiating the organic molecules using a sufficient UV radiation dose. Thereby, the organic molecules are destroyed and the fluorescence is reduced to values below the metrological detection limit. However, this measure has no lasting effect since additional organic molecules accumulate over time. Further studies have also shown that barium sulfate coatings fluoresce less than synthetic coatings. In addition, the reduction in fluorescence in case of barium sulfate coatings has a longer lasting effect. For this reason, contrary to the previous guidelines, barium sulfate is recommended as a coating material for short wavelength UV measurement applications. Due to the particularly low dose of binders in Gigahertz-Optik’s ODP97 barium sulfate coating, the topic of long-term stability under permanent UV irradiation is not critical since barium sulfate itself is UV-stable.

With the TFUV10-V01, Gigahertz-Optik offers a UV spectroradiometer for measuring the absolute radiation power of UVA, UVB and UVC LEDs in the spectral range from 200 nm to 550 nm. The instrument offers all features and functions required for this application:


Integrating sphere for radiant flux measurement

The spectroradiometer’s entrance optic for total radiant flux measurement is a 100mm diameter integrating sphere with an input aperture of 25.4 mm diameter. A sturdy port frame allows attachment of LED assemblies or the optional 2-Pi calibration standard. This standard allows the in-situ recalibration of the system by the user. The auxiliary lamp is used to compensate for the self-absorption effects caused by devices under test. The sphere coating is pre-aged using UV radiation which increases the long-term stability of the barium sulfate coating towards UV radiation and reduces fluorescence effects to below the detection limit of the measuring device. For permanent suppression of fluorescence, a UV lamp is additionally present which effectively and permanently suppresses the excitation of the fluorescence by UVC radiation below 250 nm.


High resolution spectroradiometer

The BTS2048-UV-2-F is part of the BTS2048 series, which has proven itself in a wide variety of demanding applications ranging from high-speed LED binning to the measurement of outdoor solar radiation. Such applications require the highest precision as well as reliability for continuous use. The spectroradiometer, which is attached directly to the integrating sphere without using a light guide, offers a spectral range from 200 nm to 550 nm with a spectral resolution of 1 nm. Remote-controlled optical filters increase the dynamic range. The pixels of the temperature-stabilized CCD can be synchronously set to zero. This feature offers the possibility to measure single pulses within a pulse chain.


Stray light and dark signals

Stray light and dark level offsets have a significant influence on the measurement results of UV spectroradiometers employing CCD or CMOS array sensors.

  • Stray light incorrectly illuminates pixels thereby producing measurement signals that do not result from the desired wavelength.  If stray light is not suitably suppressed, it becomes impossible to separate the LED’s real spectrum from stray light effects. The BTS2048-UV-2-F spectroradiometers offer innovative stray light suppression through integrated light traps and optical filters.
  • The influence of dark signals results from varying operating temperatures of the measuring instrument and adapted integration times for specific irradiance levels. The dark signal shutter of the BTS2048-UV-2-F spectroradiometer is used to measure the dark signal which is used for the dark signal correction of future measurements.

Data logger

The BTS2048-UV-2-F supports a high measurement repetition frequency due to its high sensitivity, powerful electronics and its fast interfaces (USB 2.0 and Ethernet). It also serves as a data logger for recording temporal signal changes.


A complete package

The TFUV10-V01 delivers a complete package for its intended use. The setting of the measuring parameters and evaluation of the measured data are fully supported by the included software. Furthermore, it allows the control of the system’s auxiliary lamp. It is also possible to use the software for system re-calibrations by the user if one of the optional calibration light sources available from Gigahertz-Optik is used.


Factory calibrations and ISO/IEC/EN 17025 testing measurements

The measurement laboratory of Gigahertz-Optik offers high quality and traceability of their TFUV10-V01 factory calibrations. Factory calibrations are handled in Gigahertz-Optik’s calibration laboratory using the same quality management procedure which applies to NMI accredited test measurements. NMI accredited testing measurements with an ISO/IEC/EN 17025 testing certificate are optionally available.

TFUV10-V01 during the measurement of BN-LDSF-2P.

General
Short description

Measurement system for spectral radiant power of UV LEDs in a wavelength range from 200 nm to 550 nm. Complete measuring system with integrating sphere, spectroradiometer BTS2048-UV-2-F, software and calibration.

Main features

High quality spectroradiometer of the BTS2048 series with high sensitivity and high spectral resolution as well as a fast data processing rate. 100 mm diameter integrating sphere, UV pre-aged for long-term stability and lowest aptitude to emit fluorescence when excited by UV radiation.

Measurement ranges

Spectral range from 200 nm to 550 nm. Spectral resolution 1 nm.

typical applications

UV LED manufacturers, UV LED luminaire manufacturers. Research institutes in the fields of semiconductors and with deep UV applcations. Metrologic institutes with demand for testing setups.

Calibration

Calibration of spectral radiant power sensitivity from 200 nm to 550 nm with detailed factory certificate including retraceability. Optional ISO/IEC/EN 17025 testing certificate of the NMI accredited test laboratory.

Spectral Detector
spectral range

(200 - 550) nm

Integration Time

2 μs - 60 s

Optical Bandwidth

1 nm

Number of pixels

2048

Chip

hochsensitiver Back-thinned CCD Chip, einstufig gekühlt (1TEC)

General

see all detailed specifications on the product page of BTS2048-UV-2

Type Description File-Type Download
Datasheet TFUV10 PDF
UMPF-1.0-HL Portframe Technical drawing of UMPF-1.0-HL Portframe
TFUV10-V01 Technical Drawing of TFUV10-V01 PDF
Produktname Product Image Description Show product
Article-Nr Modell Description Request / Compare
Product
15310312 TFUV10-V01

Measurement system for spectral radiant power from 200 nm to 550 nm.

15311205 TFUV10-V02

Measurement system for spectral radiant power from 200 nm to 430 nm.

Re-calibration
15310313 K-TFUV10-Phi-S

Re-calibration spectral radiant power from 200 nm to 550 nm of an TFUV10-V01 measurement setup. Calibration certificate.

Accessories
15307523 BN-LDSF-2P

Calibration standard, leads with banana type plugs, power supply for the fan, carry case, calibration certificate.

15298594 BN-LHSF-2P-20-100

Calibration standard, 100 W lamp, power supply for the fan, calibration certificate.