This Product is discontinued, please contact us if you are still interested. An alternative is e.g:

Several guidelines and standards for UV hazard risk assessment exist today. Most often the ACGIH/ICNIRP spectral weighting function for the acutely harmful effects of UV is referenced for irradiance measurements (American Conference of Governmental Industrial Hygienists (ACGIH) and the International Commission on Non-Ionising Radiation Protection (ICNIRP). The XD-9506-4 ACGIH detector is designed to meet EM 6 regulations. It’s two sensor design offers best possible isolation between the UV-Aeff and UV-Beff sensors while providing the closest spectral match to the ICNIRP/ACGIH spectral effectiveness function possible.

Traceable calibration

Calibration of the detector ICNIRP (W/m²) responsivity is performed by the Gigahertz-Optik GmbH calibration laboratory for optical radiation measurements quantities. As with all light detectors supplied by Gigahertz-Optik calibration of absolute detector responsivity as well as detector individual measured relative spectral responsivity data is included.

Recommended Optometer

X11

XD-9506-4 - Typical Spectral Responsivity
Specification
spectral responsivity

Ch 1: Actinic UV-AB (320 - 400) nm

Ch 2: Actinic UV-BC (250 - 320) nm

 

typical responsivity

UV-BC: 1.5 nA/(W/m²)

UV-A: 4 µA/(W/m²)

Max. signal current

UV-BC: 50 µA

UV-A: 1 mA

Input optics

8 mm Ø Cosine diffuser

Miscellaneous
temperature range

(5 - 40) °C

Cable Length

2 m

Type Description File-Type Download
Datasheet XD-9506 PDF
Produktname Product Image Description Show product
X1 Four-channel USB optometer designed for mobile use.

Features: Compact device for use with all photometric, radiometric, colorimetric, plant-physiologic and photo-biologic measurement heads from Gigahertz-Optik. USB interface. Battery operation or power supply USB.
X1-RM Optometer in 3HE housing for use in 19” racks.

Features: Its USB and RS232 remote interface and two additional RS232 device interfaces make the device highly flexible when it comes to system integration. Its four signal inputs enable use with all photometric, radiometric, colorimetric, plant-physiologic and photo-biologic measurement heads from Gigahertz-Optik.
X1-PCB Optometer module.

Feature: The X1 optometer is available as a printed circuit board either with or without a housing and is suited for applications that do not require a keyboard or display. Four signal inputs enable connection with all measuring heads from Gigahertz-Optik.
X1-PCBC Optometer module.

Feature: The X1 optometer is available as a printed circuit board either with or without a housing and is suited for applications that do not require a keyboard or display. Four signal inputs enable connection with all measuring heads from Gigahertz-Optik.
Article-Nr Modell Description Request / Compare
Re-calibration
15300571 K-UV-SR

Calibration of relative spectral responsivity from 250 nm to 550nm

15296109 XD-9506 (-4 Connector)

Detector, Calibration Certificate

15300672 K-XD9506-I

Recalibration with Calibration Certificate

Product tags