This Product is discontinued, please contact us if you are still interested. An alternative is e.g:

High Speed Digital Data Sampler for Light Pulse Form Analysis 

The TR-9600 optometers are designed for pulse locus analysis of frequency modulated and single pulse flashing light sources. 


Complete Analysis of Pulse Form and Pulse Data

  • pulse locus shape
  • peak power in absolute light measurement units
  • pulse width and pulse half width
  • single pulse energy
  • pulse repetition rate

100 ns or 1 µs rise time amplifier 

The TR-9600 analog signal amplifier offers a rise time of 1 µs (TR-9600-1) or 100 ns (TR-9600-2 *). The gain of the current to voltage amplifier is in ten steps for best adjustable signal to noise ratio.


10 mega sample / second ADC

A high speed analog to digital converter (ADC) digitizes the analog signal with up to a 10Msample/s sample rate for high time resolution measurements. Its 12 Bit resolution is higher than that 8Bit of typical digital oscilloscopes.


High-speed transient recorder with 100 ns sampling rate and pre-trigger function 

The digital data is stored in a high-speed storage medium which is designed as transient recorder allowing a sampling rate of up to 10 Mega samples per second or one sample every 100 ns. The pre-trigger function of the transient recorder enables measurement data to be stored before the triggered event. Two million of samples can be stored in memory.


Remote operation via RS232 or IEEE488 and trigger I/O interface 

The unit is capable of remote control operation via RS232 and IEEE488 interface. BNC connectors are available for trigger signal input and output or for use with external devices with TTL signal capability.


Software

Windows based software is supplied with the TR-9600 which provides all necessary functions to do remote control pulse-shape measurements and analysis via RS232 or IEEE488 interface.


Measurement Range Specifications with Light Detectors

The measurement range of the TR-9600 optometer combined with light detector is calculated by the measurement range specification of the optometer and the responsivity of the detector head as follows:

Example: Irradiance detector with typical responsivity of 3 nA/(W/cm2):

  • Maximal measurable irradiance (Range 0): 2 mA / 3 nA/(W/cm2) = 6,666,666 W/cm** 
  • Noise equivalent irradiance (Range 9): 10 mV = 0.3 nA = 10 W/cm2 
  • Minimal measurable irradiance (Range 0): 10 W/cm² * 50 (by user specified signal to noise ration) = 500 W/cm²

Limited Dynamic Range and Capacitance Limitation

Because of the wide bandwidth of the TR-9600 devices the noise level is higher than that of standard optometers which limits the dynamic range of the instrument. As a result TR-9600 detectors must be carefully selected according to the signal intensity to be measured.

The detector capacitance effected by the photodiode and the cable must be considered to avoid distortion of the pulse locus. To reduce these effects the recommended cable length for large size detectors for use with TR-9600 is 0.2 meters. This is even a greater concern with the TR-9600-2. Please consult the factory for technical support.


* The TR-9600-2 with 100 ns rise amplifier limits the freedom in selecting of the light detector because of capacitance matching as well as in the available dynamic range because of increased noise level by the wide bandwidth. Please contact therefore the factory to discuss your application.

** The maximum measurable irradiation value may also be limited by thermal radiation, intense UV radiation or other application dependent parameters which must be considered by the end-user.

TR-9600
Pulse Analysis
General
Short description

The TR-9600 Optometer has been specially developed as a data collector for the analysis of single pulses, pulse cycles or frequency modulated signals.

typical applications

Pulse locus analysis of frequency modulated and single pulse flashing light sources

Calibration

Calibration and comparison of the current responsivity in each of amplifier range.

Measurement range

1 µs rise time amplifier: 10 (1 mA/V – 30 nA/V)

100 ns rise time amplifier: 4 (300 µA/V  10 µA/V)

Main features

pulse locus shape, peak power in absolute light measurement units (depending on the detector), pulse width and pulse half width, single pulse energy, pulse repetition rate

Specification
Display

LEDs:

POWER: Device switched on, DC-power supply existing
CONNECTED: Remote-Commands received
ERROR: Error, DC-power supply low

Analog input

BNC-Socket (max. ± 5 mA, max. ± 5 V)

Bias

-12 V to + 12 V, 10 mA, Shielding BNC-Socket

Trigger

External trigger Input (TTL/CMOS, positive edge or level)

Trigger

Trigger Output (CMOS, 4 mA)

Trigger

Trigger Hysterese: ca. 50 mV

Measurement range

10 (three-step,1 mA/V – 30 nA/V)

ADC

12 Bit, max. 10 M Samples/s

sampling rate

5 Ranges (10 M Samples/s – 1 k Sample/s)

Data logger

2 M Samples (4 MByte)

Range Specifications

TR-9600-1: 10 (2.000 mA to 3.000 nA) manual or autorange

  Gain Max. Range Rise Time
(10% – 90%)
Bandwidth Noise
(p-p)
Error
(20ºC)
AR0 1 mA/V ± 2 mA 1 µs 333 kHz 2 mV < 1% ± 2 mV
AR1 300 µA/V ± 600 µA 1 µs 333 kHz 3 mV < 1% ± 2 mV
AR2 100 µA/V ± 200 µA 1 µs 333 kHz 2 mV < 1% ± 2 mV
AR3 30 µA/V ± 60 µA 1 µs 333 kHz 3 mV < 1% ± 2 mV
AR4 10 µA/V ± 20 µA 1 µs 333 kHz 2 mV < 1% ± 2 mV
AR5 3 µA/V ± 6 µA 1 µs 333 kHz 4 mV < 1% ± 2 mV
AR6 1 µA/V ± 2 µA 3 µs 111 kHz 4 mV < 1% ± 2 mV
AR7 300 nA/V ± 600 nA 3 µs 111 kHz 10 mV < 1% ± 2 mV
AR8 100 nA/V ± 200 nA 30 µs 11 kHz 4 mV < 1% ± 2 mV
AR9 30 nA/V ± 60 nA 30 µs 11 kHz 10 mV < 1% ± 2 mV
Version
TR-9600-1

1 µs version

TR-9600-2

100 ns version

Miscellaneous
Power Supply

(6 - 8) V / 1.2 A, DC power plug 5.5 mm / 2.5 mm / 10 mm
AC/DC power supply (7V/1.5 A)

temperature range

5 °C – 40 °C

Interface

RS232 (75 - 57600Baud, 8 Data-Bit's, 1 Stop-Bit, no Parity)
DSUB9-Socket:

PIN 2: TxD
PIN 3: RxD
PIN 5: GND
PIN 1,4,6 connected
PIN 7,8 connected


IEEE488 (488-1978, AH1, SH1, L4, T4)

Dimensions

w = 28,0 cm

d = 25,2 cm

h = 7,8 cm

Weight

2000 g (without battery)

Warranty

12 months

Type Description File-Type Download
Datasheet TR-9600 PDF
Produktname Product Image Description Show product
VL-3701 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, f2 ≤ 1.5 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3702 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, f2 ≤ 3 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3704 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 20pA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3705 Detector head for the measurement of scotopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 0.2nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

PD-9310A High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration certificate

PD-9310B High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration

PD-9310B-N Very high sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 28nA/lx, no diffuser, for the usage with optometers and amplifiers, calibration

VL-1101 + UMPA-0.5-11-RD Detector head Module detector head for the measurement of photopic illuminance in Lux [lx].

Features: UMPA adapter for usage with integrating spheres, for the usage with optometers and amplifiers, calibration certificate

VL-3701 with SRT-M37-L Detector head to measure the photopic illuminance in lx and the luminance in cd/m².

Features: front lenses with 1°, 2° or 5° viewing angle, for the usage with Optometers and amplifiers, calibration certificate

PD-9310 with SRT-M37-L High sensitive detector head to measure the photopic luminance in cd/m².

Features: front lense for 1°, 2°, 5° or 10° viewing angle, for the usage with Optometers and amplifiers, calibration certificate

VL-1101 Photometric detector head with VL-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc. For use with optometers and signal amplifiers

PD-9304 Universal detector head for LASER power, illuminance and 400-1100 nm iiradiance.

Features: Si-photodiode with 1 cm², exchange able filters and cosine diffuor, for the usage with optometers and signal amplifiers

RW-3701 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-500nm (BLUE), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3702 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 700-800nm (RED), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3703 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-800nm (VIS), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3704 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 800-1000nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
RW-3705 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-1000nm (VISNIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3708 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 950-1700nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3701 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3702 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-315nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3703 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 200/250-280nm (UV-C), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3710 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 320-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3711 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-320nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3716 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 305-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3717 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), low cross talk from radiation > 400 nm, cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3719 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 250-400nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3720 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 240-320nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3718 Detector head for the measurement of high irradiance of UV-C 254nm radiation in W/m².

Features: pre-aging for long time stability, cosine field-of-view, metal shielded cable, for the usage with optometers and amplifiers, calibration certificate.

PD-11 series Detector head with DP-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc, Si, SiLP, InGaAs, SiC, GaP photodiodes, for use with optometers and signal amplifiers

UV-3706 Detector head to measure irradiance W/m2 in Bilirubin phototherapy.

Features: Bilirubin actinic responsivity, cosine field-of-view, for usewith optometers, calibration certificate.

UV-3711-308 Detector head for the measurement of irradiance of 308nm Eximer Lasers in W/m².

Features: flat spectral responsivity besite 308nm. cosine field-of-view, dose measurement in conjuntion with P-9710-2 optometer, calibration certificate

UV-3709 Detector for Blue-light hazard measurements.

Features: Single-cell detector, BLH actinic irradiance, for the use with optometer, calibration certificate

UV-3725 Detector for the measurement of UV-C 254 nm irradiance in air disinfection applications.

Features: wide dynamic range for UV hazard and effective irradiance, wide angle cosine F.O.V. for straylight measurements, for the usage with optometers, calibration certificate.
RW-37 with SRT-M37-L Detector heads to measure the irradiance in W/m² and the radiance in W/(m²*sr).

Features: front lenses with 1°, 2° or 5° viewing angle, for the usage with Optometers and amplifiers, calibration certificate
S-SDK-TR9600 Software Development Kit for TR9600 variants.
S-TR9600 Application software for the TR9600 variants.
RCH-102 Detector head for high intensity irradiation in UV or blue light curing processes.

Features: Separate light integrator and detector with rigid fiber coupling, (320-460)nm UVABLUE responsivity, wide viewing angle, for the usage with optometers and amplifiers, calibration certificate.
ISD-5-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 50mm dia, BaSO4 coating, 12.5mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 100 mm dia, BaSO4 coating, 15 mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

TD-11VL01 Photometric, temperature stabilized detector with DP-11 mount.

Features: for use with integrating spheres, Operation temperature 50 °C, for use with optometers and signal amplifiers

LP-9901 Detector head to measure LASER radiant power in W and LASER irradiance in W/m².

Features: Low height detector with 7mm dia active area, 400 to 1100nm, for the usage with optometers and signal amplifiers, calibration certificate.

ISD-3P-Si Integrating sphere detector for Laser power in W.

Features: 30 mm dia, 5 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-5P-Si Integrating sphere detector for Laser power in W.

Features: 50 mm dia, 10 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-Si Integrating sphere detector for Laser power in W.

Features: 100mm dia, 15mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

LDM-9811 Detector for Blue-light and retinal thermal hazard of expanded light sources.

Features: 1.7mrad, 11mrad and 100mrad field-of-view. view finder, focusable, for the usage with Optometers, calibration certificate.
PRW-0505 Integrating sphere detector for radiant power in W of 2π spot sources.

Features: 400 - 1100 nm radiometric responsivity, 50 mm dia integrating sphere, 12.5 mm dia measurement port, for the usage with optometers and signal amplifiers, Calibration certificate
ISD-3P-IGA Integrating sphere detector with InGaAs photodiode and 30 mm sphere for Laser power in W.

Features: 800 nm - 1800 nm spectral responsivity, 5 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate
ISD-5P-SiUV Integrating sphere detector with UV-enhanced Si photodiode and 50 mm sphere for Laser power in W

Features: 250 nm - 1100 nm spectral responsivity, 10 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate

ISD-5P-IGA Integrating sphere detector with InGaAs photodiode and 50 mm sphere for Laser power in W.

Features: 800 nm - 1800 nm spectral responsivity, 10 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate

Article-Nr Modell Description Request / Compare
Product
15295243 TR-9600-1

Meter, with power supply, software and manual

15295277 TR-9600-2

Meter, with power supply software and manual

Options
- Light Detectors

Please check the light detector datasheets or the tab configurable with for specification and purchasing information

Software
15298648 S-SDK-TR9600

Software Development Kit for the implementation of the TR9600 or variants into custom made software

15298646 S-TR9600

User software for TR9600 and variants.

Accessories
15295220 BHO-02

Hard case for meter and accessories