P-9710-1 – Multi-functional light meter for high-quality photometric measurements

The P-9710-1 optometer is a high-quality device designed by Gigahertz-Optik for mobile use. Its signal amplifier provides a wide dynamic range for the detectors’ signal currents ranging up to 2 mA. The noise-equivalent current signal is as low as 0.1 pA. The device has eight gain ranges that can be manually or automatically selected for adjustment to match the signal levels.

Variable integration time

One of the characteristic properties of the P-9710-1 optometer is the short integration time of the input amplifier that only ranges around 100 µs. This enables reduction of signal noises through multiple measurements with averaging. The 100 µs integration time also makes it possible to perform many other measurement functions:

  • Adjustable integration times in CW measurement mode
  • CW measurement with noise proportion (peak-to-peak)
  • Peak value measurement
  • Pulse energy measurements of short light pulses
  • Dose measurement
  • Fast data logger for up to 12000 measurement values

Photometer, radiometer, UV-meter, laser-intensity-meter

For use as optometer the P-9710-1 can be combined with all offered detector heads intended for photometric and radiometric measurements. The plugged detector head determines the photometric or radiometric measuring size as well as the spectral range of the measurement. Gigahertz-Optik offers a wide range of detector heads for various measuring tasks. The optometers’ signal input amplifiers offer a very wide dynamic range, in order to adjust the devices to detectors with various sensitivities and optical radiation intensities. The calibration data of the detector heads are saved in the calibration data plug of the detector heads.

Simple and safe detector exchange

Upon attachment of a detector head with a calibration data plug onto the P-9710-1 optometer, its calibration data as well as the device settings are automatically applied. This ensures that improper operation that might result from exchange of measurement heads is basically ruled out.

Remote operation

The P-9710-1 optometer can be connected to a PC via its RS232 interface.

Light measurements and light analysis

The P-9710-1 optometer has eighteen different measurement modes that can easily be selected using a user-friendly menu hence enabling its multifunctional application.

 

P-9710-2 – Optometer for dose measurement of short pulses using the pulse stretching method

Unlike the P-9710-1, the P-9710-2 optometer has a 20 ms standard time constant in all eight gain ranges. Thus light pulses shorter than 1 ns are stretched to 20 ms. The pulse energy of the stretched pulse can then be measured precisely with the 100 µs sample rate of the optometer.

 

P-9710-4 – multi-functional light meter for synchronized measurement of short light pulses

Similar to the optometer P-9710-2, the P-9710-4 also has a 20ms standard time constant in all eight gain levels. Besides the manual start of via keyboard, the dose measurement can also be started remotely via the trigger input.

The P-9710-4 offers a trigger input for a remotely controlled triggering of measurement
The optometers P-9710 provide connections for power supply, detector heads with calibration data plug and a RS232 interface (from left to right).
General
Measured Quantity

Ampere (absolutely calibrated), light measurement units depending on the used detector head. Ratio as a percentage, logarithmic and as a factor

P-9710-1

High-quality optometer for mobile use. Trans-impedance amplifier with alterable rise times in 8 gain levels. Mainly used in CW measurements.

P-9710-2

Equivalent to the optometer P-9710-1, but with a time constant of 20ms in all eight gain levels. Mainly used for measuring the pulse engergy of single pulses.

P-9710-4

Equivalent to the optometer P-9710-2, with an additional trigger input for a synchronized triggering of the pulse energy measurement.

Specification
Display

Alpha-numerical LCD display, 2 rows each with 16 characters, height of the characters 5 mm, LED backlight can be switched on and off.

Detector interface

9pin DSUB

Analog output

Output voltage corresponding to the input current of the signal amplifier. Connection plug: TRIAD01 5pin (Tyco)

Signal amplifier

Trans-impedance amplifier with voltage amplifier connected in series. A total of eight gain levels with both manual and automatic selection of the measurement range.

CW integration time

100 μs  5.9999 s

Pulse integration time

10 ms  199.99 s

Offset correction

Exceeding correction range

Parameter adjustment

Via RS232 or using control panel on the front side. Settings are permanently saved (EEPROM). Ten different settings can be saved

Calibration

Saved in the detector terminal (EEPROM). Tables for calibration values of the spectral responsivity with up to 250 values are supported. Interpolation between two wavelength points are also possible. Manual input of a calibration value via the key pad is also possible.

Data logger

Max. 12288 entries, permanently saved in flash memory

Key pad

10 buttons, menu-guided

Measurement ranges

 8 (2.000 mA to 0.1 pA) manual or autorange

range

(A/V)

range

max.

slew rate

(10 - 90%)

variant -1

slew rate

(10 - 90%)

variant -2 / -4

gain error *) ± offset error

(@ 20 °C)

gain (A/V)
analog output

1x10-3

±2.000mA

2 ms

20 ms

0.2% ± 0.001 mA

1x10-3

1x10-4

±200.0 µA

2 ms

20 ms

0.2% ± 0.1 µA

1x10-3

1x10-5

±20.00 µA

3 ms

20 ms

0.2% ± 0.01 µA

1x10-5

1x10-6

±2.000 µA

3 ms

20 ms

0.2% ± 0.001 µA

1x10-5

1x10-7

±200.0 nA

4 ms

20 ms

0.2% ± 0.1 nA

1x10-7

1x10-8

±20.00 nA

4 ms

20 ms

0.2% ± 0.01 nA

1x10-7

1x10-9

±2.000 nA

10 ms

20 ms

0.5% ± 2 pA

1x10-9

1x10-10

±200.0 pA

10 ms

20 ms

0.5% ± 2 pA

1x10-9

Miscellaneous
Power Supply

Rechargeable battery, battery capacity: 3,7 V, 1400 mAh

Rechargeable battery, battery capacity: 6 V, 500 Ah (Version before 2017)

 

Operation time: 5 hours with display backlight switched on, 12 hours with display backlight switched off

5.5 mm / 2.5 mm / 9.5 mm socket for plug-in power supply unit

Interface

RS232 (9600 baud, 8 data bits, 1 stop bit, no parity)
Plug: TRIAD01 (Tyco)

temperature range

Operating: (5 to 40) °C
Storage: (-10 °C to 50) °C

Dimensions

195 mm x 100 mm x 40 mm / 500 g

Accessories
Power Supply

D, USA or GB

Type Description File-Type Download
Datasheet P-9710 PDF
Produktname Product Image Description Show product
VL-3701 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, f2 ≤ 1.5 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3702 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, f2 ≤ 3 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3704 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 20pA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3705 Detector head for the measurement of scotopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 0.2nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

PD-9310A High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration certificate

PD-9310B High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration

PD-9310B-N Very high sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 28nA/lx, no diffuser, for the usage with optometers and amplifiers, calibration

LDM-9810 Detector head to measure the photopic spot luminance in cd/m².

Features: selectable 20', 1° and 6° viewing angles, view finder, focus able achromatic lens, for the usage with Optometers and amplifiers, calibration certificate.

LDM-9901 Detector head to measure the photopic spot luminance in cd/m².

Features: 1.1° viewing angle, simple to use, very low stray light, for the usage with Optometers and amplifiers, calibration certificate.

PD-9310 with SRT-M37-L High sensitive detector head to measure the photopic luminance in cd/m².

Features: front lense for 1°, 2°, 5° or 10° viewing angle, for the usage with Optometers and amplifiers, calibration certificate

ISD-5-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 50mm dia, BaSO4 coating, 12.5mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 100 mm dia, BaSO4 coating, 15 mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-15P-VL Integrating sphere detector for luminous flux (lm) of 2π sources.

Features: 150mm dia, synthetic coating, 38.1mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

MD-37-SU100-VL Photometric detector head with M30x1 mount.

Features: modular detector for use MD-37, SRT and other accessories. For use with optometers and signal amplifiers

MD-37-SU100-VLS Scotopic detector head with M30x1 mount.

Features: modular detector for use MD-37, SRT and other accessories. For use with optometers and signal amplifiers

PD-9304 Universal detector head for LASER power, illuminance and 400-1100 nm iiradiance.

Features: Si-photodiode with 1 cm², exchange able filters and cosine diffuor, for the usage with optometers and signal amplifiers

RW-3701 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-500nm (BLUE), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3702 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 700-800nm (RED), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3703 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-800nm (VIS), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3704 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 800-1000nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
RW-3705 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-1000nm (VISNIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3708 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 950-1700nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RCH-102 Detector head for high intensity irradiation in UV or blue light curing processes.

Features: Separate light integrator and detector with rigid fiber coupling, (320-460)nm UVABLUE responsivity, wide viewing angle, for the usage with optometers and amplifiers, calibration certificate.
ISD-5-VISNIR Integrating sphere detector for radiant power in W of 2π sources.

Features: 400 nm - 1100 nm radiometric responsivity, 50 mm dia, 12.5 mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate
ISD-0.8-SiLP Integrating sphere detector with short rise time for Laser power in W.

Features: 8 mm dia, ODM98 coating, 400 nm - 1100 nm responsivity, optional SMA Adapter, for the usage with fast optometers and signal amplifiers, Calibration certificate

ISD-1.6 Si Integrating sphere detector for Laser power in W.

Features: 16mm dia, 3mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-3P-Si Integrating sphere detector for Laser power in W.

Features: 30 mm dia, 5 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-5-Si Integrating sphere detector for Laser power in W.

Features: 50 mm dia, 12.5 mm dia port, BaSO4 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-5P-Si Integrating sphere detector for Laser power in W.

Features: 50 mm dia, 10 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-Si Integrating sphere detector for Laser power in W.

Features: 100mm dia, 15mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-15-Si Integrating sphere detector for Laser power in W.

Features: 150mm dia, 38.1mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-30 Integrating sphere detector for Laser power in W.

Features: 300mm dia, 101.6mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

PD-11 series Detector head with DP-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc, Si, SiLP, InGaAs, SiC, GaP photodiodes, for use with optometers and signal amplifiers

MD-37 series Detector head with M30x1 mount.

Features: modular detector for use MD-37, SRT and other accessories, Si, SiLP, InGaAs, SiC, GaP photodiodes, for use with optometers and signal amplifiers

PS-3701 Detector head for plant growth.

Features: 400-700nm PAR actinic irradiance, cosine field-of-view, for use with optometers, calibration certificate.

PS-3702 Detector head for plant growth.

Features: 320-500nm phototropism actinic irradiance, for use with optometers, calibration certificate.

PS-3703 Detector head for plant growth.

Features: 590-900nm photomorphogenesis actinic irradiance, for use with optometers, calibration certificate.

S-P9710 Application software for P9710 variants.
VL-1101 + UMPA-0.5-11-RD Detector head Module detector head for the measurement of photopic illuminance in Lux [lx].

Features: UMPA adapter for usage with integrating spheres, for the usage with optometers and amplifiers, calibration certificate

VL-3701 with SRT-M37-L Detector head to measure the photopic illuminance in lx and the luminance in cd/m².

Features: front lenses with 1°, 2° or 5° viewing angle, for the usage with Optometers and amplifiers, calibration certificate

VL-1101 Photometric detector head with VL-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc. For use with optometers and signal amplifiers

TD-11VL01 Photometric, temperature stabilized detector with DP-11 mount.

Features: for use with integrating spheres, Operation temperature 50 °C, for use with optometers and signal amplifiers

UV-3701 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3702 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-315nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3703 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 200/250-280nm (UV-C), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3710 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 320-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3711 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-320nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3716 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 305-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3717 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), low cross talk from radiation > 400 nm, cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3719 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 250-400nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3720 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 240-320nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3718 Detector head for the measurement of high irradiance of UV-C 254nm radiation in W/m².

Features: pre-aging for long time stability, cosine field-of-view, metal shielded cable, for the usage with optometers and amplifiers, calibration certificate.

LP-9901 Detector head to measure LASER radiant power in W and LASER irradiance in W/m².

Features: Low height detector with 7mm dia active area, 400 to 1100nm, for the usage with optometers and signal amplifiers, calibration certificate.

UV-3706 Detector head to measure irradiance W/m2 in Bilirubin phototherapy.

Features: Bilirubin actinic responsivity, cosine field-of-view, for usewith optometers, calibration certificate.

UV-3711-308 Detector head for the measurement of irradiance of 308nm Eximer Lasers in W/m².

Features: flat spectral responsivity besite 308nm. cosine field-of-view, dose measurement in conjuntion with P-9710-2 optometer, calibration certificate

LDM-9811 Detector for Blue-light and retinal thermal hazard of expanded light sources.

Features: 1.7mrad, 11mrad and 100mrad field-of-view. view finder, focusable, for the usage with Optometers, calibration certificate.
UV-3709 Detector for Blue-light hazard measurements.

Features: Single-cell detector, BLH actinic irradiance, for the use with optometer, calibration certificate

UV-3725 Detector for the measurement of UV-C 254 nm irradiance in air disinfection applications.

Features: wide dynamic range for UV hazard and effective irradiance, wide angle cosine F.O.V. for straylight measurements, for the usage with optometers, calibration certificate.
RW-37 with SRT-M37-L Detector heads to measure the irradiance in W/m² and the radiance in W/(m²*sr).

Features: front lenses with 1°, 2° or 5° viewing angle, for the usage with Optometers and amplifiers, calibration certificate
ISS-5P Integrating sphere source for spectral radiance.

Features: Compact size with 20mm dia light port. In-line baffle. Synthetic ODM98 coating. Halogen lamp. Calibration of the spectral radiance 380-1100nm. Calibration certificate.

ISS-5P-SR-FS Integrating sphere source for use as transfer standard for spectral radiance in fluorescence spectroscopie.

Features: Low spectral radiance. Compact size with 20mm dia light port. Synthetic ODM98 coating. lamp. Calibration of the spectral radiance 380-1700nm. Calibration certificate.
S-SDK-P9710 Software Development Kit for the P9710 variants.
Article-Nr Modell Description Request / Compare
Product
15295217 P-9710-1

Measurement device, power adapter (D, USA or GB), handbook

15295575 P-9710-2

Meter, power supply (D, USA or GB), manual

15295576 P-9710-4

Meter, power supply (D, USA or GB), manual

Options
Light detectors

See "Configurable with" for all detector heads suitable with the P-9710. More information about each detector head are given at it's product page.

Software
15298278 S-P9710

User software for P9710 and variants.

15298231 S-SDK-P9710

Software Development Kit for the implementation of the P9710 or variants into custom made software

Accessories
100001 P-9710Z-01

RS232 adapter cable

100371 P-9710Z-02

Relay card with RS232 interface

100867 P-9710Z-04

Connection plugs kit for combined RS232 and analog output socket

100385 P-9710Z-1S/2B

Adapter cable for detectors with BNC (type 1) plugs

100010 BHO-01

Hard-top casing for the measurement device and accessories

100002 BHO-02

Hard-top casing for the measurement device and accessories