High efficient two channel optometer

Available in two versions the P-2000 optometers are highly efficient dual-channel instruments designed for multipurpose use in most photometric and radiometric applications including pulsed light measurement. Its compact size, four line blue back-lit display and unlimited detector interchange, functions and features characterize the P-2000 as all purpose laboratory grade instrument.


Safe Detector Head exchange 

A unique feature of the P-2000 is its detector head calibration data connector which stores all data pertaining to a light detector head including model and serial number and calibration data. When connected to the meter, this data is automatically transmitted and the light meter is ready to go. Combined with one or more of the wide range of available light detectors the P-2000 can be configured as a high level photometer, UV-A, B and C radiometer, laser power meter, PAR meter plus many other configurations.


Seventeen different measurement modes 

Along with the ability to use the P-2000 optometers with an unlimited range of light detectors, seventeen measurement modes enable the user to specify light sources in different ways.


CW and pulse energy measurement 

The P-2000-1 features a fast sample rate ADC with a variable integration time from 0.1 to 6 seconds in CW mode. The P-2000-2 model is designed for pulse energy measurement of single pulses or pulse chains down to µs pulse lengths.


Manual and remote control operation 

The RS232 and IEEE488 interfaces allows remote control operation. The meter is DC low voltage operated with external AC power supply.


Measurement range specifications with Light Detectors

The measurement range of optometer combined with light detector is calculated by the measurement range specification of the optometer and the responsivity of the detector head as follows:

  • Offset signal
    = Maximum Resolution = meter current offset signal / detector sensitivity
    Sample: 0.1 pA (0.1E-12 A) / 3 nA/(mW/cm²) (irradiance detector) = 0.33 nW/cm²
  • Minimum measureable irradiation
    = offset signal * signal to noise ratio factor
    Sample: 0.33 nW/cm² * 50 = 17 nW/cm²
  • Maximum measureable irradiation *
    = max. signal current detector / detector sensitivity
    Sample: 1 mA (10E-3 A) / 3 nA/(mW/cm²) = 333333 W/cm²
  • Display range
    = Offset signal to maximum measureable signal
    Sample: 0.33 nW/cm² to 333333 W/cm²
  • Measurement range
    = min measureable irradiation to maximum measureable irradiation
    Sample: 17 nW/cm² to 333333 W/cm²

*) The maximum measureable irradiation value may also be limited by thermal radiation, intense UV radiation or other application dependent parameters which must be considered by the end-user.


 


P-2000Z-02

Relay board for P-2000 to control external accessories (e. g. yellow, green and red indication lamps for low-ok-high indications in binning processes). Solid state relays controlled via P-2000 RS232 interface. Low-ok-value setting in the CW level set mode of the P-2000.

 

P-2000 with Optional Integrating Sphere Detector
P-2000 with Relay Board (Option)
P-2000Z-01 RS232 cable required for remote control operation
Relay board for P-2000 to control external accessories
Adapter cable to connect detectors with BNC connectors (-1 type) to the 9PIN SUBD socket
Product
Detector interface

2x 9pin DSUB

Analog output

Output voltage corresponding to detector input current (Ri = 10k), connector: BNC

Inputs

Amperes

optical units corresponding to calibration data

factor, percent, log (depending on mode)

Measurement range

8 ranges (2.000 mA to 0.1 pA) manual or auto range

CW integration time

100 µs – 5.9999 s

Measurement time

Puls: 10 ms – 199.99 s

Offset correction

Correction range transcending

Parameter adjustment

Remote control or front panel keys (menu), adjusted values permanently stored (EEPROM)

Calibration information

Stored in the detector connector (EEPROM)
Manual calibration factor (keys)
Max. 250 calibration table entries + interpolation between the entries

Logger memory

Max. 5734 entries for each channel, permanently stored in flash memory

Interface

RS232: 9600 Baud, 8 Data Bit, 1 Stop Bit, No Parity, Connector DSUB 9pins,
female IEEE488: AH1, SH1, L4, T4

Rise time

Input Amplifier Slew Rate: 20 ms

Measurement range
range
(A/V)
range
max.
slew rate
(10 – 90%)
P-2000-1
slew rate
(10 – 90%)
P-2000-2
gain error *)
± offset error
(at 20 °C)
gain (A/V)
analogue output
1x10-3 ± 2.000 mA 2 ms 20 ms 0.2% ± 0.001 mA 1x10-3
1x10-4 ± 200.0 µA 2 ms 20 ms 0.2% ± 0.1 µA 1x10-3
1x10-5 ± 20.00 µA 3 ms 20 ms 0.2% ± 0.01 µA 1x10-5
1x10-6 ± 2.000 µA 3 ms 20 ms 0.2% ± 0.001 µA 1x10-5
1x10-7 ± 200.0 nA 4 ms 20 ms 0.2% ± 0.1 nA 1x10-7
1x10-8 ± 20.00 nA 4 ms 20 ms 0.2% ± 0.01 nA 1x10-7
1x10-9 ± 2.000 nA 10 ms 20 ms 0.5% ± 2 pA 1x10-9
1x10-10 ± 200.0 pA 10 ms 20 ms 0.5% ± 2 pA 1x10-9


*) Current calibration of each range by use of a precise current source with DAkks calibration

Interface
DSUB 9pin female Function
1 input current (detector current)
2 GND
3 connection detection (has to be connected to GND by the external detector, this is the case by Gigahertz-Optik detectors)
4 GND
5 SCL (I2C, eeprom)
6 GND
7 GND
8 +5V Output (Ri = 100 Ohm)
9 SDA (I2C, eeprom)
Output Interfaces
Pin (TRIAD01 female) Function
1 Analog Out (Rin = 10k)
2 TxD (has to be connected to RxD of ext. PC)
3 RxD (has to be connected to TxD of ext. PC)
4 GND
5 GND
Measurement modes
Menu Item Submenu Item Function
1. Mode CW displays the measurement respective of any offset and calibration factors programmed
CW Maximum displays the highest detected reading
CW Minimum displays the lowest detected reading
Peak Maximum shows the maximum detected peak level
Peak Minimum displays the minimum detected peak level
Peak to Peak displays the difference between the detected peak maximum and peak minimum level
I-Effective enables the measurement of pulsed light signals with evaluation of the effective intensity according to the form-factor method defined by Schmidt-Clausen
Pulse Energy enables the measurement of short and single light pulses with direct display of exposure for the pulse duration, considering the respectively measured (radiometric) quantity
Relative (%) displays the measurement as percentage of a reference value
Relative (Log) displays the measurement in dB or dBm in relation to a reference value
Relative (Fact.) displays the measurement referred to a reference value
Relative (CCT) displays the CCT value calculated out of the current ratio of channel 1 and channel 2
CW Level Check compares actual CW readings with previously defined limit values and indicates status
Dose (C) accumulates the single readings and displays the result as exposure for measured quantity
Logger stores the single readings, which are taken in predefined intervals up to a number of 5734 for each channel
Remote RS232 enables the instrument to be controlled by a computer via the built in RS232 interface
Remote IEEE488 enables the instrument to be controlled by a computer via the built in IEEE488 interface
2. Range  selects the range 1 to 8 or switches into the autorange mode (9)
3. Detector Detector information if a detector with calibration data connector is attached to the instrument, optionally programmed calibration factors can be selected
Ampere shows the reading in ampere units
Manual allows entering a calibration factor manually
4. Offset
performs an automatic offset adjustment
5. Reference
sets a reference value, used in different display modes
6. Setup Integration time sets the integration time (100 µs to 6 s)
Zero Adjust Mode sets the mode for the internal zero adjustment
Pulse Meas.-Time needs to be set for the measuring modes I-Effective and Pulse Energy. It defines the time for collecting readings and shall be a little longer than the expected pulse width.
IF Time Constant sets the time constant C in the Schmidt-Clausen formula
Pulse Offset determines the method of offset compensation (stray light) for the measuring modes I-Effective and Pulse Energy
CW Level Check sets the lower / higher limit value required in the CW Level Check measuring mode
Dose Run Time sets the maximum time duration for exposure measurement
Dose Maximum sets the max. dose level for the exposure measurement mode
Dose Relay Control enables / disables external relay board control via RS232 for mode dose
Logger Time sets the sample time interval for the data logger mode
Display Digits adjusts the number of displayed digits (4, 5, or automatic)
Default Init resets all instrument settings to the factory default settings
Channel selects how the displayed channel
IEEE488 Address adjusts the device address for IEEE488
Synchronisation selects how the measurement time period is adapted to the input signal
7. Info Logger data displays the recorded logger data
CW Level Min. sets the lower limit value required in the CW Level Check measuring mode
CW Level Max. sets the upper limit value required in the CW Level Check measuring mode
Default Init resets all instrument settings to the factory default settings
Synchronization selects how the measurement time period is adapted to the input signal
Substitution enables / disables self absorptin error compensation (substitution correction)
Code Number sets a four digit lock-out access code
Configuration Default Init – sets factory default settings. Save Config – stores all current settings. Load Config. – loads setting selection (0-9) for Save Config.
8. Info Battery Status displays the battery charge status as percentage value
Logger data displays the recorded logger data
Version

P-2000-1: Input Amplifier Slew Rate: 2ms  – 10 ms

P-2000-2: Input Amplifier Slew Rate: 20 ms

Miscellaneous
Power Supply

(6.5 – 7.5) VDC / 300 mA, Plug 5.5 / 2.5 mm / 10 mm

Display

Alphanumeric LCD display, 4 rows x 20 chars, character height 5 mm, LED-illumination

Front panel control

10 keys, menu system

temperature range

Operation: (5 to 40) °C
Storage: (-10 to 50) °C

Dimensions

230 mm x 215 mm x 115 mm

Power Supply

230 AC / 7.5 VDC / 500 mA

Warranty

12 months

Weight

800 g

Type Description File-Type Download
Datasheet P-2000 PDF
Produktname Product Image Description Show product
VL-3701 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, f2 ≤ 1.5 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3702 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, f2 ≤ 3 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3704 Detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 20pA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

VL-3705 Detector head for the measurement of scotopic illuminance in Lux [lx].

Features: f1 ≤ 5 %, f2 ≤ 3 %, 0.2nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate

PD-9310A High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration certificate

PD-9310B High sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 6 %, 2.8nA/lx, 20mm diffuser, for the usage with optometers and amplifiers, calibration

PD-9310B-N Very high sensitive detector head for the measurement of photopic illuminance in Lux [lx].

Features: f1 ≤ 3 %, 28nA/lx, no diffuser, for the usage with optometers and amplifiers, calibration

LDM-9810 Detector head to measure the photopic spot luminance in cd/m².

Features: selectable 20', 1° and 6° viewing angles, view finder, focus able achromatic lens, for the usage with Optometers and amplifiers, calibration certificate.

LDM-9901 Detector head to measure the photopic spot luminance in cd/m².

Features: 1.1° viewing angle, simple to use, very low stray light, for the usage with Optometers and amplifiers, calibration certificate.

VL-1101 Photometric detector head with VL-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc. For use with optometers and signal amplifiers

ISD-5-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 50mm dia, BaSO4 coating, 12.5mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-VL Integrating sphere detector for luminous flux (lm) of 2π spot sources.

Features: 100 mm dia, BaSO4 coating, 15 mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-15P-VL Integrating sphere detector for luminous flux (lm) of 2π sources.

Features: 150mm dia, synthetic coating, 38.1mm dia port, for the usage with optometers and signal amplifiers, Calibration certificate

TD-11VL01 Photometric, temperature stabilized detector with DP-11 mount.

Features: for use with integrating spheres, Operation temperature 50 °C, for use with optometers and signal amplifiers

PD-9304 Universal detector head for LASER power, illuminance and 400-1100 nm iiradiance.

Features: Si-photodiode with 1 cm², exchange able filters and cosine diffuor, for the usage with optometers and signal amplifiers

RW-3701 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-500nm (BLUE), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3702 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 700-800nm (RED), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3703 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-800nm (VIS), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3704 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 800-1000nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
RW-3705 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 400-1000nm (VISNIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

RW-3708 Detector head for the measurement of irradiance in W/m².

Features: spectral responsivity from 950-1700nm (NIR), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3701 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3702 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-315nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3703 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 200/250-280nm (UV-C), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3710 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 320-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3711 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 280-320nm (UV-B), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3716 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 305-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.
UV-3717 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 315-400nm (UV-A), low cross talk from radiation > 400 nm, cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3719 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 250-400nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3720 Detector head for the measurement of irradiance of UV radiation in W/m².

Features: spectral responsivity from 240-320nm (UV), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate.

UV-3718 Detector head for the measurement of high irradiance of UV-C 254nm radiation in W/m².

Features: pre-aging for long time stability, cosine field-of-view, metal shielded cable, for the usage with optometers and amplifiers, calibration certificate.

RCH-102 Detector head for high intensity irradiation in UV or blue light curing processes.

Features: Separate light integrator and detector with rigid fiber coupling, (320-460)nm UVABLUE responsivity, wide viewing angle, for the usage with optometers and amplifiers, calibration certificate.
LP-9901 Detector head to measure LASER radiant power in W and LASER irradiance in W/m².

Features: Low height detector with 7mm dia active area, 400 to 1100nm, for the usage with optometers and signal amplifiers, calibration certificate.

ISD-3P-Si Integrating sphere detector for Laser power in W.

Features: 30 mm dia, 5 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-5-Si Integrating sphere detector for Laser power in W.

Features: 50 mm dia, 12.5 mm dia port, BaSO4 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-5P-Si Integrating sphere detector for Laser power in W.

Features: 50 mm dia, 10 mm dia port, synthetic ODM98 coating, 400 nm - 1100 nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-10-Si Integrating sphere detector for Laser power in W.

Features: 100mm dia, 15mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

ISD-30 Integrating sphere detector for Laser power in W.

Features: 300mm dia, 101.6mm dia port, BaSO4 coating, 400-1100nm responsivity, for the usage with optometers and signal amplifiers, Calibration certificate

PD-11 series Detector head with DP-11 mount.

Features: modular detector for use with integrating spheres, front lenses etc, Si, SiLP, InGaAs, SiC, GaP photodiodes, for use with optometers and signal amplifiers

UV-3706 Detector head to measure irradiance W/m2 in Bilirubin phototherapy.

Features: Bilirubin actinic responsivity, cosine field-of-view, for usewith optometers, calibration certificate.

UV-3711-308 Detector head for the measurement of irradiance of 308nm Eximer Lasers in W/m².

Features: flat spectral responsivity besite 308nm. cosine field-of-view, dose measurement in conjuntion with P-9710-2 optometer, calibration certificate

UV-3709 Detector for Blue-light hazard measurements.

Features: Single-cell detector, BLH actinic irradiance, for the use with optometer, calibration certificate

UV-3725 Detector for the measurement of UV-C 254 nm irradiance in air disinfection applications.

Features: wide dynamic range for UV hazard and effective irradiance, wide angle cosine F.O.V. for straylight measurements, for the usage with optometers, calibration certificate.
PS-3701 Detector head for plant growth.

Features: 400-700nm PAR actinic irradiance, cosine field-of-view, for use with optometers, calibration certificate.

PS-3702 Detector head for plant growth.

Features: 320-500nm phototropism actinic irradiance, for use with optometers, calibration certificate.

PS-3703 Detector head for plant growth.

Features: 590-900nm photomorphogenesis actinic irradiance, for use with optometers, calibration certificate.

RW-37 with SRT-M37-L Detector heads to measure the irradiance in W/m² and the radiance in W/(m²*sr).

Features: front lenses with 1°, 2° or 5° viewing angle, for the usage with Optometers and amplifiers, calibration certificate
PRW-0505 Integrating sphere detector for radiant power in W of 2π spot sources.

Features: 400 - 1100 nm radiometric responsivity, 50 mm dia integrating sphere, 12.5 mm dia measurement port, for the usage with optometers and signal amplifiers, Calibration certificate
ISD-3P-IGA Integrating sphere detector with InGaAs photodiode and 30 mm sphere for Laser power in W.

Features: 800 nm - 1800 nm spectral responsivity, 5 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate
ISD-5P-SiUV Integrating sphere detector with UV-enhanced Si photodiode and 50 mm sphere for Laser power in W

Features: 250 nm - 1100 nm spectral responsivity, 10 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate

ISD-5P-IGA Integrating sphere detector with InGaAs photodiode and 50 mm sphere for Laser power in W.

Features: 800 nm - 1800 nm spectral responsivity, 10 mm dia measurement port, synthetic ODM98 coating, optional SMA Adapter, for the usage with Optometer and signal amplifiers, calibration certificate

S-SDK-P2000 Software Development Kit for P2000 and variants.
Article-Nr Modell Description Request / Compare
Product
15295970 P-2000-1

Meter, power supply, manual

15297598 P-2000-2

Meter, power supply, manual

Software
15298227 S-SDK-P2000

Software Development Kit for the implementation of a P2000 or variants into custom made software

Accessories
15296020 P-2000Z-01

RS232 adapter cable

15296034 P-2000Z-02

Relay board with RS232 interface

15296235 P-9710Z-1S/2S

Adapter cable to connect detector heads with BNC (-1) type connector

Product tags