Contact
Events
This Product is discontinued, please contact us if you are still interested. An alternative is e.g:

A more economically priced calibration than KLD-S1 with slightly higher calibration uncertainties called the KLW spectral irradiance calibration is available which includes a factory certificate. The device data, calibration procedure, calibration uncertainty, environmental conditions as well as the traceability data back to the national standards laboratory are confirmed in the certificate. 

The lamp power range is unlimited. Wavelength range is from 250 nm to 2500 nm. 

 

 

Wavelength Dependent Calibration Uncertainty
Measured Quantity

Spectral Irradiance in W/(m² nm) of Optical Radiation Sources

Note

Power range >100 W. Measurement uncertainty may increase depending on lamp type

KLW-S1
Wavelength nm Relative Measurement Uncertainty
λ=250 ± 12 %
λ=260 ± 8 %
270 ≤ λ <400 ± 4.5 %
400 ≤ λ<800 ± 3.5 %
800 ≤ λ <2000 ± 5.5 %
2000 ≤ λ ≤ 2500 ± 8.5 %
Calibration Steps

10 nm from 250 to 300 nm; 20 nm from 320 to 800 nm; 50 nm from 850 to 2500 nm

Type Description File-Type Download
Datasheet KLW-S1 PDF
Produktname Product Image Description Show product
BN-LH250 Calibration lamp for spectral irradiance and illuminance.

Features: 250 W halogen lamp with stable and therefore long-lasting, helical structure. Lamp holder, crosshair. Calibration in spectral irradiance, 250 nm - 2500 nm, and/or illuminance, with factory or DAkkS certificate, optional power supply.
BN-9101 Calibration standard lamp for spectral irradiance.

Features: 1000W tungsten halogen lamp. Burn-in certificate. Stable filament. Horizontal calibration from 250-2500nm with factory or DAkkS certificate.
BN-0001 Calibration standard lamp for spectral irradiance.

Features: 1000W tungsten halogen lamp. Burn-in certificate. Stable filament. Vertical calibration from 250-2500nm with factory or DAkkS certificate.

Article-Nr Modell Description Request / Compare
Calibration
KLW-S1-01

Spectral Irradiance calibration within the wavelength range 250 to 400 nm with factory certificate

KLW-S1-02

Spectral Irradiance calibration within the wavelength range 400 to 1100 nm with factory certificate

KLW-S1-03

Spectral Irradiance calibration within the wavelength range 1100 to 2500 nm with factory certificate

KLW-S1-04

Spectral Irradiance calibration within the wavelength range 250 to 1100 nm with factory certificate

KLW-S1-05

Spectral Irradiance calibration within the wavelength range 400 to 2500 nm with factory certificate

KLW-S1-06

Spectral Irradiance calibration within the wavelength range 250 to 2500 nm with factory certificate

KLW-S1-RP

Set-up charge for non-Gigahertz-Optik sources

Product tags