Dieses Produkt ist abgekündigt, bitte kontaktieren sie uns bei Interesse. Eine Alternative ist z.B.:

Calibration of spectral sensitivity with a factory calibration certificate is offered for the wavelength range from 200 nm to 1800 nm in 10 nm increments. The calibration is done by com- parison to an PTB or NIST traceable transfer standard detector. A factory calibration certificate is supplied listing detector data, measurement set-up, traceability and calibration uncertainty.

Wavelength Dependent Calibration Uncertainty
Measurement Quantity

Spectral Radiant Power Sensitivity

Light Spot Diameter

3 - 7 mm

Relative Measurement Uncertainty

200 - 240 nm ± 7 %
250 - 370 nm ± 4.5 %
380 - 400 nm ± 3.0 %
410 - 900 nm ± 3.0 %
910 - 1070 nm ± 4.5 %
1080 - 1100 nm ± 5 %
1110 - 1800 nm ± 5.5 %

KDW-S1
200 to 400 in 10 nm steps 1 nW/cm² ≤ E ≤ 10 μW/cm²
18 °C ≤ t ≤ 28 °C
1 nm ≤ Δλ ≤ 11 nm
400 to 1100 in 10 nm steps 0.1 μW/cm² ≤ E ≤ 100 μW/cm² 
18 °C ≤ t ≤ 28 °C
1 nm ≤ Δλ ≤ 11 nm
1100 to 1800 in 10 nm steps 0.1 μW/cm² ≤ E ≤ 100 μW/cm² 
18 °C ≤ t ≤ 28 °C
2 nm ≤ Δλ ≤ 22 nm
Typ Beschreibung Datei-Typ Download
Datenblatt KDW-S1 PDF
Produktname Produktbild Beschreibung Zum Produkt
Artikel-Nr Modell Beschreibung Merken / Vergleichen
Kalibrierung
KDW-S1-01

Spectral radiant power sensitivity calibration in 10 nm steps from 200 to 400 nm with factory certificate

KDW-S1-02

Spectral radiant power sensitivity calibration in 10 nm steps from 350 to 1100 nm with factory certificate

KDW-S1-03

Spectral radiant power sensitivity calibration in 10 nm steps from 1000 to 1800 nm with factory certificate

KDW-S1-04

Spectral radiant power sensitivity calibration in 10 nm steps from 200 to 1100 nm with factory certificate

KDW-S1-05

Spectral radiant power sensitivity calibration in 10 nm steps from 350 to 1800 nm with factory certificate

KDW-S1-06

Spectral radiant power sensitivity calibration in 10 nm steps from 200 to 1800 nm with factory certificate

KDW-S1-07

Spectral radiant power sensitivity calibration at one wavelength within 200 to 400 nm with factory certificate

KDW-S1-08

Additional wavelength to KDW-S1-07

KDW-S1-09

Spectral radiant power sensitivity calibration at one wavelength within 350 to 1100 nm with factory certificate

KDW-S1-10

Additional wavelength to KDW-S1-09

KDW-S1-11

Spectral radiant power sensitivity calibration at one wavelength within 1000 to 1800 nm with factory certificate

KDW-S1-12

Additional wavelength to KDW-S1-11

KDW-S1-RP

Set-up charge for non-Gigahertz-Optik detectors