Dieses Produkt ist abgekündigt, bitte kontaktieren sie uns bei Interesse. Eine Alternative ist z.B.:

Besides standard detector responsivity calibration services, Gigahertz-Optik offers a field-of-view calibration service for light detectors. The FOV calibration is offered for all irradiance and illuminance detectors employing diffuser windows for a cosine corrected measurement geometry. The calibration is done with a light source at a distance from the detector. The test detector is mounted onto a goniometric rotary table. It’s diffuser surface is aligned to the axis of the rotary table. The signal as a function of the viewing angle is measured and compared with the ideal cosine spatial function. The measured characteristic as well as the calculated f2 uncertainty is stated in the factory calibration certificate.       

   

 

 

Typ Beschreibung Datei-Typ Download
Datenblatt K-FOV PDF
Produktname Produktbild Beschreibung Zum Produkt
Artikel-Nr Modell Beschreibung Merken / Vergleichen
Kalibrierung
15300577 K-FOV-01

Field-of-View calibration of irradiance or illuminance light detectors with f2 uncertainty calculation. Factory calibration certificate

15300493 K-FOV-02

Field-of-View calibration of two cells irradiance or illuminance light detectors with f2 uncertainty calculation. Factory calibration certificate