<< zurück

ISMP-50-R-V01



Measurement of Reflectance

The ISMP-50-R-V01 is designed to perform total reflectance as well as the diffuse reflectance component measurement of flat surface test samples. The regular reflectance component can be calculated by the difference between the measured values.


8/d Measurement Geometry with Substitution Correction

Its design conforms to DIN 5036 T3 for 8/d measurement geometry with 8° sample illumination and integrating sphere to measure reflected light with an almost hemispherical acceptance angle. Measurement errors caused by the self absorption effect of the test sample can be compensated using the substitution correction port.


50cm / 12 in Diameter Integrating Sphere

The complete instrument includes an integrating sphere with accessory components, light source and supporting electronics. The integrating sphere and light source are mounted to a bench top stand for stability and to prevent misalignment during transportation and in use. The sphere itself is 50cm in diameter and includes a measurement port, auxiliary port, illumination port (+8° orientated to sample) and regular reflectance output port (-8° orientated to the sample). The sphere diameter to port size relation is in accordance with DIN 5036 T3 for traceable measurement results.


8° Incident Sample Illumination

A collimated light source employing a halogen tungsten lamp operated at the color temperature of 2896 K (standard illuminant A) is used for precise sample illumination. The lamp is operated in constant current mode for best stability.


Spring Loaded Sample Holder

Both the measurement and auxiliary port include a spring loaded sample holder for positioning the test sample at the reference plane on the port. The electronics include a sophisticated lightmeter/controller with photometric detector mounted onto the sphere and light source power supply. The meter/controller supports the substitution correction measurement routine, the calibration routine and displays the light reflectance measurement values in percent as well as remote control of the power supply.


Diffuse Spectral Reflectance Calibration Standards

For measurement applications involving samples with diffuse reflection characteristics Gigahertz-Optik offers a diffuse reflection calibration standard model BN-R98-D2C-PP. Calibration of the diffuse calibration standard is traceable to NRC. For non-diffuse test samples (flat surface mirrors) the BN-RM95-50C-PP mirror type reflectance calibration standard is available with calibration traceable to PTB. Both standards are supplied in a housing with locating pins on the rear side which fit into holes on the sample holder.

Product Pictures



















  schließen

Specifications



Electronics Specifications:
Optometer/Controller Identical to model X1-RM. Firmware set for reflectance measurements with substitution correction and calibration. Measurement value display in %. Remote On/Off switch control of lamp power supply.
Detector interface 9pin MDSM9 socket, 4 inputs
Remote interface USB V1.1 (HID Device)
RS232 19200Baud (Slave Mode) for LPS-100 remote control
Display LCD graphic display 97x32 pixel
Display area 14.3mm x 35.8mm
Switchable LED-backlight
Text display 4 rows each 14 characters
Front panel control 3 buttons, menu system
Temperature Operating: 5 to 40°C
Storage: -10 to 50° C
Lamp Power Supply Identical to model LPS-100-RM. Firmware set constant current mode and remote control by X1-RM. Lamp On/Off ramp mode activated.
Output power 100W / 0 - 10A / max. 13.5V
Output connector Phoenix MSTB25/4GF (5.00)
Resolution of output current 0.2mA
Temperature stability +-0.3mA / °K
Long term stability +-1mA (60 min after switch on of output current, constant environment) for 8 hours)
LED green Blinking slow: DC input voltage present
Blinking fast: Current ramp active
Continuous: Output switched on
LED red Continuous: error condition
Remote interface RS232 / RS485: 19200 Bd, 8D, 1S, N
Temperature range Operation: 10 - 30°C (with free air flow), Storage: 5 - 50°C
Electronic Device Set-up in BTH-19 bench-top housing with BTH-19-Z03 power supply
Operation Power 120 to 230 VAC, 50 to 60Hz
Size 3/4 19” 3HE Bench-top Housing
Integrating Sphere Specifications:
Integrating Sphere 50 cm diameter. ODP97 barium sulfate coating
Measurement port 50mm diameter with knife edge design
Auxiliary port 50mm diameter with knife edge design
Sample holder Spring loaded holder and port cover. Orientation holes for calibration standard
-8° port 25mm diameter. 8° orientated to the sample surface. Port plug and alternative light trap with 10mm aperture. Slide & Fix holder
+8° port 50mm diameter, 8° orientated to the sample surface
Measurement Detector Identical to PD-9310A-4
Photometric matching f1’ uncertainty <6% (DIN 5052 T7 class B)
Measurement signal ca. 20.3nA with BN-R98-D2 reflectance standard and –8° port closed
Light source Collimated beam with 8° orientation to the test sample surface. Aligned by mirror to the center of the –8° port
Color temperature 2856K +/-20K (Standard illuminant A)
Lamp LH-100-UV, 100W/12V quartz halogen lamp. Exchangeable by user
Cooling Active airflow cooling of the lamp socket by two fans
Connection to LPS Two laboratory type leads with banana type plugs
Light Beam ca. 18mm at sample, ca. 9mm at –8° port
Warm up-time ca. 10min
ISMP-50-R-V01 Measurement Device Specifications:
Measurement Range 0.5% to 100% with 0.01% resolution of the measurement value
Dynamic range 200 (dark signal 0.5% measured with open -8° and low scattering mirror at the measurement port.
Measurement uncertainty ± 0.5% plus calibration standard uncertainty
Measurement reproducibility ± 0,1*[1-log10rg]%; rg = regular/hemispherical
CE conformity The device conforms with the requirements set out in the council directive on the approximation of the laws of the member states relating to electromagnetic compatibility (89/336/EEC). For the evaluation regarding electromagnetic compatibility, the following standards were applied: EN 61326-1 - Class B
Rohs Compatible
Warranty 12 month
Manual English



Calibration Standard Uncertainty:
BN-R98-D2C-PP ± 0.6% (k=2), diffuse reflecting type (ODM98)
BN-RM95-50C-PP ± 0.5% (k=2), regular reflecting type (plan, very low scattering mirror)

  schließen

Purchasing Information



Model Item No.: Description
ISMP-50-R-V01 102731 Light reflectance measurement instrument with integrating sphere and control electronics, connection cables, manual (English)
BN-RM95-50C-PP 102704 Diffuse reflectance standard with calibration certificate
BN-R98-D2C-PP 102697-1 Specular reflectance standard with calibration certificate