TR-9600-1
TR-9600-2



High Speed Digital Data Sampler for Light Pulse Form Analysis

The TR-9600 optometers are designed for pulse locus analysis of frequency modulated and single pulse flashing light sources.

Complete Analysis of Pulse Form and Pulse Data
  • pulse locus shape
  • peak power in absolute light measurement units
  • pulse width and pulse half width
  • single pulse energy
  • pulse repetition rate
100ns and 1µs rise time amplifier

The TR-9600 analog signal amplifier offers a rise time of 1µs (TR-9600-1) or 100ns (TR-9600-2*). The gain of the current to voltage amplifier is in ten steps for best adjustable signal to noise ratio.

10 mega sample / second ADC

A high speed analog to digital converter (ADC) digitizes the analog signal with up to a 10Msample/s sample rate for high time resolution measurements. Its 12Bit resolution is higher than that 8Bit of typical digital oscilloscopes.

High-speed transient recorder with 100ns sampling rate and pre-trigger function

The digital data is stored in a high-speed storage medium which is designed as transient recorder allowing a sampling rate of up to 10 Mega samples per second or one sample every 100 ns. The pre-trigger function of the transient recorder enables measurement data to be stored before the triggered event. Two million of samples can be stored in memory.

Remote operation via RS232 or IEEE488 and trigger I/O interface

The unit is capable of remote control operation via RS232 and IEEE488 interface. BNC connectors are available for trigger signal input and output or for use with external devices with TTL signal capability.

Software

Windows based software is supplied with the TR-9600 which provides all necessary functions to do remote control pulse-shape measurements and analysis via RS232 interface.

* The TR-9600-2 with 100ns rise amplifier limits the freedom in selecting of the light detector because of capacitance matching as well as in the available dynamic range because of increased noise level by the wide bandwidth. Please contact therefore the factory to discuss your application.






TR-9600 Optometer



Pulse Locus Measurement



Pulse Analysis

OS-TR9600 Software

Specifications



General Specifications Model TR-9600-1, TR-9600-2:
Power 6 - 8V / 1.2A, DC power plug 5.5 mm/ 2.5 mm/ 10 mm
AC/DC power supply (7V/1.5 A)
Temperature Range 5 - 40 °C
Display-LED's POWER: Device switched on, DC- power supply existing
CONNECTED: Remote-Commands received
ERROR: Error, DC-power supply low
Analogue-Input BNC-Socket (max. +-5mA, max. +-5V)
Bias Voltage Detector 12V to + 12V, 10mA, Shielding BNC-Socket
Trigger-Input External trigger Input (TTL/CMOS, positive edge or level)
Trigger-Output Trigger Output (CMOS, 4mA)
Trigger-Hysteresis ca. 50mV
Measurement Range 10 ( three-step,1mA/V - 30nA/V)
ADC 12Bit, max. 10MSamples/s
Sampling-Rate 5 Ranges (10MSamples/s - 1kSample/s)
Memory for Measurement Data 2MSamples (4MByte)
Remote Interface RS232 (75 - 57600Baud, 8 Data-Bit's, 1 Stop-Bit, no Parity)
DSUB9-Socket: PIN 2: TxD
PIN 3: RxD
PIN 5: GND
PIN 1,4,6 connected
PIN 7,8 connected
IEEE488 (488-1978, AH1, SH1, L4, T4)
Size w= 28,0 cm, d = 25,2 cm, h=7,8 cm
Weight 2000g (without battery)
CE conformity The device conforms with the requirements set out in the council directive on the approximation of the laws of the member states relating to electromagnetic compatibility (89/336/EEC). For the evaluation regarding electromagnetic compatibility, the following standards were applied: EN 61326-1 - Class B
Rohs Compatible
Warranty 12 month
Range Specifications TR-9600-1:
Remote Gain Max. Range Slew-Rate
(10 - 90%)
Noise
(p-p)
Error
(20ºC)
AR0 1mA/V +-2mA 1µs 2mV <1% +-2mV
AR1 300µA/V +-600µA 1µs 3mV <1% +-2mV
AR2 100µA/V +-200µA 1µs 2mV <1% +-2mV
AR3 30µA/V +-60µA 1µs 3mV <1% +-2mV
AR4 10µA/V +-20µA 1µs 2mV <1% +-2mV
AR5 3µA/V +-6µA 1µs 4mV <1% +-2mV
AR6 1µA/V +-2µA 3 µs 4mV <1% +-2mV
AR7 300nA/V +-600nA 3 µs 10mV <1% +-2mV
AR8 100nA/V +-200nA 30 µs 4mV <1% +-2mV
AR9 30nA/V +-60nA 30 µs 10mV <1% +-2mV

  close

Measurement Range Specifications with Light Detectors:



Measurement Range Specifications with Light Detectors:

The measurement range of the TR-9600 optometer combined with light detector is calculated by the measurement range specification of the optometer and the responsivity of the detector head as follows:

Example: Irradiance detector with typical responsivity of 3nA/(W/cm2):

Maximal measurable irradiance (Range 0): 2mA / 3nA/(W/cm2) = 6,666,666W/cm2 *)

Noise equivalent irradiance (Range 9): 10mV = 0.3nA = 10W/cm2

Minimal measurable irradiance (Range 0): 10W/cm² x 50 (by user specified signal to noise ration) = 500W/cm²

*)The maximum measureable irradiation value may also be limited by thermal radiation, intense UV radiation or other application dependent parameters which must be considered by the end-user.



Limited Dynamic Range and Capacitance Limitation:

Because of the wide bandwidth of the TR-9600 devices the noise level is higher than that of standard optometers which limits the dynamic range of the instrument. As a result TR-9600 detectors must be carefully selected according to the signal intensity to be measured.

The detector capacitance effected by the photodiode and the cable must be considered to avoid distortion of the pulse locus. To reduce these effects the recommended cable length for large size detectors for use with TR-9600 is 0.2 meters. This is even a greater concern with the TR-9600-2. Please consult the factory for technical support.

  close

Light Detector to configure TR-9600 as Photometer



The TR-9600 Optometers in combination with one or more Gigahertz-Optik photometric light detectors or one luminous colour detector forms an universal photometer for demanding light measurement applications:

  • Illuminance measurement (lx)
  • Luminous Intensity measurement (cd)
  • Luminance measurement (cd/m²)
  • Luminous Flux measurement (lm)
  • Dose (e.g. lx*s) measurement


Model Function Spectral Range
nm
Detection
Area Size
mm
F.O.V. Sphere
Dia
mm
Typ.
Responsivity
Max
Signal
Compact Size Illuminance Detectors
VL-1101-1 Photopic; f1≤5% 380-780 1.1 Ø - - 0.3nA/lx1) 1mA
Photopic & Scotopic Illuminance Detectors
VL-3701-1 Photopic; f1≤3% 380-780 7 Ø Cosine - 0.5nA/lx 1mA
VL-3702-1 Photopic; f1≤5% 380-780 7 Ø Cosine - 0.5nA/lx 1mA
VL-3704-1 Photopic; f1≤5% 380-780 7 Ø Cosine - 20pA/lx 1mA
VL-3705-1 Photopic; f1≤5% 380-780 7 Ø Cosine - 0.5nA/lx 1mA
PD-9310A-1 Photopic; f1≤3% 380-780 30 Ø Cosine - 2.8nA/lx 1mA
PD-9310A-N-1 Photopic; f1≤3% 380-780 11,28 Ø No Diffuser - 28nA/lx 1mA
PD-9310B-1 Photopic; f1≤6% 380-780 30 Ø Cosine - 2.8nA/lx 1mA
PD-9310B-N-1 Photopic; f1≤6% 380-780 11,28 Ø No Diffuser - 28nA/lx 1mA
Temperature Controlled Illuminance Detectors
TD-11VL01-1 Photopic; f1≤3% 380-780 1.1 Ø - - 0.3nA/lx1) 1mA
Luminance Detectors
LDM-9810 &
PD-16VL01-1
Photopic; f1≤3% 380-780 50 Ø 20‘

- 0.0035nA/(cd*m²)
0.033nA/(cd*m²)
0.9nA/(cd*m²)
1mA
LDM-9901-1 Photopic; f1≤5% 380-780 22 Ø 1.1° - 25pA/(cd*m²) 1mA
VL-3701-1
SRT-M37-L1
Photopic; f1≤5% 380-780 22.4 Ø - 1.2pA/(cd*m²) 1mA
VL-3701-1
SRT-M37-L2
Photopic; f1≤5% 380-780 22.4 Ø - 5pA/(cd*m²) 1mA
VL-3701-1
SRT-M37-L5
Photopic; f1≤5% 380-780 22.4 Ø - 30pA/(cd*m²) 1mA
PD-9310A-1*) &
SRT-M37-L10
Photopic; f1≤3% 380-780 22.4 Ø 10° - 1.7nA/(cd*m²) 1mA
Luminous Flux Detectors
ISD-5-1 Photopic; f1≤6% 380-780 12.5Ø - 50 10nA/(lm) 1mA
ISD-10-1 Photopic; f1≤6% 380-780 15Ø - 100 4nA/(lm) 1mA
ISD-15-1 Photopic; f1≤6% 380-780 25Ø - 150 4A/(lm) 1mA
ISD-30-1 Photopic; f1≤6% 380-780 75Ø - 300 1nA/(lm) 1mA
ISD-50HF**) Photopic; f1≤6% 380-780 **) - 500 tbc A/(lm) 1mA
ISD-100HF**) Photopic; f1≤6% 380-780 **) - 1000 tbc A/(lm) 1mA
*) VL-1101-1 Detector **) Test Sample Lamp in the Sphere Centre

  close

Light Detector to configure TR-9600 as UV-VIS-NIR Radiometer



The P-9801 Optometer in combination with one or more Gigahertz-Optik radiometric light detectors forms a universal radiometer for demanding optical radiation measurement applications:

  • Irradiance Measurement (W/m², W/cm²)
  • Radiant Intensity Measurement (W/sr)
  • Radiance Measurement (W/(sr*m²))
  • Measurement of Dose (e.g. J/m²) measurement


Model Spectral Function Spectral
Range
nm
Detection
Area Size
mm
F.O.V. Sphere
Dia
mm
Typ.
Responsivity
Max
Signal
High Sensitive Irradiance Detectors
RW-3701-1 Radiometric BLUE 400-500 11 Ø Cosine - 70nA/(W/m²) 1mA
RW-3702-1 Radiometric RED 700-800 11 Ø Cosine - 70nA/(W/m²) 1mA
RW-3703-1 Radiometric VISIBLE 400-800 11 Ø Cosine - 1.7nA/(W/m²) 1mA
RW-3704-1 Radiometric NIR 800-1000 11 Ø Cosine - 200nA/(W/m²) 1mA
RW-3705-1 Radiometric VISNIR 400-1000 11 Ø Cosine - 400nA/(W/m²) 1mA
RW-3708-1 Radiometric NIR 1000-1700 11 Ø Cosine - 40nnA/(W/m²) 1mA
High Sensitivity UV Irradiance Detectors
UV-3701-1 Radiometric UV-A 315-400 11 Ø Cosine - 4nA/(W/m²) 100µA
UV-3702-1 Radiometric UV-B 280-315 11 Ø Cosine - 1.7nA/(W/m²) 50µA
UV-3703-1 Radiometric UV-C 250-280 11 Ø Cosine - 2A/(W/m²) 50µA
UV-3710-1 Radiometric UV-A 320-400 11 Ø Cosine - 5A/(W/m²) 100µA
UV-3711-1 Radiometric UV-B 280-320 11 Ø Cosine - 1/(W/m²) 50µA
UV-3716-1 Radiometric UV-A 305-400 11 Ø Cosine - 10nA/(W/m²) 100µA
UV-3717-1 UV-A Visible Blind 325-400 11 Ø Cosine - 2.5A/(W/m²) 100µA
UV-3718-1 Narrowband UV-C 254 11 Ø Cosine - 2nA/(W/m²) 50µA
UV-3719-1 Broadband UV 250-400 11 Ø Cosine - 3nA/(W/m²) 100µA
Radiant Power Detectors
PRW-0505-1 Radiometric VIS-NIR 400-1000 12.5 Ø - 50 90µA/W 1mA
Radiance Detectors for Photobiological Actinic Radiation:
LDM-9811 with
PD-16RW05-1
Radiometric VISNIR 400-1000 50 Ø 1.75mrad
11mrad
100mrad
-
-
-
0.38nA/(W/m²sr)
12nA/(W/m²sr)
840nA/(W/m²sr)
1mA

  close

Light Detector to configure TR-9600 as UV & BLUE Light Curing Radiometer



The TR-9600 Optometer in combination with one or more of Gigahertz-Optik irradiance detectors for high intensity and high temperature radiation forms a universal radiometer for demanding optical radiation measurement in UV and BLUE light curing applications:

  • Irradiance Measurement in W/m²
  • Irradiance Dose Measurement in J/m²


Model Function Spectral
Range
nm
Detection
Area Size
mm
F.O.V. Sphere
Dia
mm
Typ.
Responsivity
Max
Signal
Irradiance Detectors with flexible fiber for ARC Lamps
RCH-002-1 UVA-BLUE 320-475 9 Ø Cosine - 0.2nA/(mW/cm²) 100µA
RCH-005-1 UVC-BLUE 250-470 9 Ø Cosine - 0.1nA/(mW/cm²) 100µA
RCH-006-1 UVC-UVA 250-400 9 Ø Cosine - 0.3nA/(mW/cm²) 50µA
RCH-008-1 UVA 365 peak 9 Ø Cosine - 0.3nA/(mW/cm²) 100µA
RCH-009-1 BLUE 430 peak 9 Ø Cosine - 0.2nA/(mW/cm²) 100µA
RCH-010-1 UVC-UVB 240-320 9 Ø Cosine - 40nA/(mW/cm²) 100µA
RCH-011-1 UVA 350-400 9 Ø Cosine - 400nA/(mW/cm²) 100µA
RCH-012-1 BLUE 400-450 9 Ø Cosine - 600nA/(mW/cm²) 100µA
Irradiance Detectors with rigid fiber for ARC Lamps
RCH-102-1 UVA-BLUE 320-475nm 320-475 9 Ø Cosine - 0.2nA/(mW/cm²) 100µA
RCH-106-1 UVC-UVA 250-400 9 Ø Cosine - 0.3nA/(mW/cm²) 50µA
RCH-108-1 UVA 365 peak 9 Ø Cosine - 0.3nA/(mW/cm²) 100µA
RCH-109-1 BLUE 430 peak 9 Ø Cosine - 0.2nA/(mW/cm²) 100µA
RCH-114-1 BLUE 405&436 9 Ø Cosine - 0.1nA/(mW/cm²) 100µA
RCH-115-1 BLUE 436 peak 9 Ø Cosine - 0.2nA/(mW/cm²) 100µA
Irradiance Detectors with rigid fiber for high power LEDs
RCH-116-1 UVA-BLUE 360-450 9 Ø Cosine - 1.8nA/(mW/cm²) 100µA

  close

Light Detector to configure TR-9600 as LASER Power and Power Density Meter



The TR-9600 Optometer in combination with one or more of Gigahertz-Optik detectors for LASER power forms a precise radiometer for demanding radiation measurement of low power LASERs:

  • Radiant Power Measurement in W from 200 to 1800nm
  • Power Density Measurement in W/m² fro 400 to 1100nm


Model Function Spectral
Range
nm
Detection
Area Size
mm
Sphere
Dia
mm
Typ.
Responsivity
Max
Signal
Radiant Power with Flat-field Detector
LP-9901-1 Silicon with ND Filter 400-1100 7 Ø - 1.3mA/W @633nm
20mA/W @900nm
0.9mA/(W/cm²) @633nm
13mA/W/cm²) @900nm
1mA
PD-9304-1 Silicon 380-1100 11.28 Ø - 400mA/W @633nm
650mA/W @900nm
400mA/(W/cm²) @633nm
650mA/(W/cm²) @900nm
1mA
Radiant Power with Integrating Sphere
ISD-3P-Si-1 Silicon 350-1100 5 Ø 30 7mA/W @633nm
10mA/W @900nm
1mA
ISD-3P-IGA-1 InGaAs 800-1100 5 Ø 30 0.1mA/W @1000nm
0.2mA/W @1300nm
1mA
ISD-5P-Si-1 Silicon 350-1100 10 Ø 30 2mA/W @633nm
2.5mA/W @900nm
1mA
ISD-5P-SiU-1 UV-Enhanced Silicon 250-1100 10 Ø 50 0.2A/W @250nm
2mA/W @633nm
2.5mA/W @900nm
1mA
ISD-5P-IGA-1 InGaAs 800-1100 10 Ø 50 0.06mA/W @1000nm
1mA/W @1300nm
1mA
ISD-10-Si-1 Silicon 350-1100 15 Ø 100 0.9mA/W @633nm
1.1mA/W @900nm
1mA

  close

Purchasing Information



Model Item No.: Description
TR-9600-1 100024 Meter, with power supply, software and manual
TR-9600-2 100054 Meter, with power supply software and manual
BHO-02 100002 Hard case for meter and accessories
Light Detectors Please check the light detector datasheets for specification and purchasing information