Light Reflectance Measurement Instrument with 8/d Geometry per DIN 5036-T3
The ISMP-50-R-V01 is designed to perform total reflectance as well as the diffuse reflectance component measurement of flat surface test samples. The regular reflectance component can be calculated by the difference between the measured values. Its principal design follows the recommendations found in DIN 5036 T3 for 8/d measurement geometry with 8° sample illumination and integrating sphere to measure re-reflected light with an almost hemispherical acceptance angle. Measurement errors caused by the self absorption effect of the test sample can be compensated using the substitution correction port.
The instrument is composed of an integrating sphere with accessory components, light source and supporting electronics. The integrating sphere and the light source are mounted to a stable bench top stand to avoid misalignment during transportation and in use. The sphere itself is 50cm in diameter and includes a measurement port, auxiliary port, illumination port (+8° orientated to sample) and regular reflectance output port (-8° orientated to the sample). The sphere diameter to port size relation is in accordance with DIN 5036 T3 for traceable measurement results. A collimated light source employing a halogen tungsten lamp operated at the color temperature of 2896 K (standard illuminant A) is used for precise sample illumination. The lamp is operated in constant current mode for best stability. Both, the measurement and auxiliary port include a spring load sample holder for positioning the test sample at the reference plane on the port. The electronics include a sophisticated lightmeter/controller with photometric detector mounted onto the sphere and light source power supply. The meter/controller supports the substitution correction measurement routine, the calibration routine and displays the light reflectance measurement values in percent as well as remote control of the power supply.
For measurement applications involving samples with diffuse reflection characteristics Gigahertz-Optik offers a diffuse reflection calibration standard model BN-R98-D2C-PP. Calibration of the diffuse calibration standard is traceable to NRC. For non-diffuse test samples (flat surface mirrors) the BN-RM95-50C-PP mirror type reflectance calibration standard is available with calibration traceable to PTB. Both standards are supplied in a housing with orientation pins on their back side which fit into holes on the sample holder.
To monitor instrument calibration consider keeping one standard as a transfer standard between Gigahertz-Optik and the end-user. This standard should be re-calibrated annually by Gigahertz-Optik and then used for recalibration of the working by the end-user. This way the end-user has a traceable in-house reference standard for periodical check of its measurement device.
ISMP-50-R-V01 measurements should be done under dimmed lighting conditions to avoid ambient light influence on the readings. Ambient light levels can be checked by making a dark signal measurement with the light meter.
Product Pictures
Specifications (Data)
Electronics Specifications:
Integrating Sphere Specifications:
ISMP-50-R-V01 Measurement Device Specifications:
Calibration Standard Uncertainty:
| Optometer/Controller | Identical to model X1-RM. Firmware set for reflectance measurements with substitution correction and calibration. Measurement value display in %. Remote On/Off switch control of lamp power supply. |
| Detector interface | 9pin MDSM9 socket, 4 inputs |
| Remote interface | USB V1.1 (HID Device) RS232 19200Baud (Slave Mode) for LPS-100 remote control |
| Display |
LCD graphic display 97x32 pixel Display area 14.3mm x 35.8mm Switchable LED-backlight Text display 4 rows each 14 characters |
| Front panel control | 3 buttons, menu system |
| Temperature |
Operating: 5 to 40°C Storage: -10 to 50° C |
| Lamp Power Supply | Identical to model LPS-100-RM. Firmware set constant current mode and remote control by X1-RM. Lamp On/Off ramp mode activated. |
| Output power | 100W / 0 - 10A / max. 13.5V |
| Output connector | Phoenix MSTB25/4GF (5.00) |
| Resolution of output current | 0.2mA |
| Temperature stability | +-0.3mA / °K |
| Long term stability | +-1mA (60 min after switch on of output current, constant environment) for 8 hours) |
| LED green |
Blinking slow: DC input voltage present Blinking fast: Current ramp active Continuous: output switched on |
| LED red | Continuous: error condition |
| Remote interface | RS232 / RS485: 19200 Bd, 8D, 1S, N |
| Temperature range | Operation: 10 - 30°C (with free air flow), Storage: 5 - 50°C |
| Electronic Device | Set-up in BTH-19 bench-top housing with BTH-19-Z03 power supply |
| Operation Power | 120 to 230 VAC, 50 to 60Hz |
| Size | 3/4 19” 3HE Bench-top Housing |
| Integrating Sphere | 50 cm diameter. ODP97 barium sulfate coating |
| Measurement port | 50mm diameter with knife edge design |
| Auxiliary port | 50mm diameter with knife edge design |
| Sample holder | Spring loaded holder and port cover. Orientation holes for calibration standard |
| -8° port | 25mm diameter. 8° orientated to the sample surface. Port plug and alternative light trap with 10mm aperture. Slide & Fix holder |
| +8° port | 50mm diameter, 8° orientated to the sample surface |
| Measurement Detector | Identical to PD-9310A-4 |
| Photometric matching | f1’ uncertainty <6% (DIN 5052 T7 class B) |
| Measurement signal | ca. 20.3nA with BN-R98-D2 reflectance standard and –8° port closed |
| Light source | Collimated beam with 8° orientation to the test sample surface. Aligned by mirror to the center of the –8° port |
| Color temperature | 2856K +/-20K (Standard illuminant A) |
| Lamp | LH-100-UV, 100W/12V quartz halogen lamp. Exchangeable by user |
| Cooling | Active airflow cooling of the lamp socket by two fans |
| Connection to LPS | Two laboratory type leads with banana type plugs |
| Light Beam | ca. 18mm at sample, ca. 9mm at –8° port |
| Warm up-time | ca. 10min |
| Measurement Range | 0.5% to 100% with 0.01% resolution of the measurement value |
| Dynamic range | 200 (dark signal 0.5% measured with open -8° and low scattering mirror at the measurement port. |
| Measurement uncertainty | ± 0.5% plus calibration standard uncertainty |
| Measurement reproducibility | ± 0,1*[1-log10rg]%; rg = regular/hemispherical |
| -CE conformity | The device conforms with the requirements set out in the council directive on the approximation of the laws of the member states relating to electromagnetic compatibility (89/336/EEC). For the evaluation regarding electromagnetic compatibility, the following standards were applied: EN 61326-1 - Class B |
| Rohs | Compatible |
| Warranty | 12 month |
| Manual | English |
| BN-R98-D2C-PP | ± 0.6% (k=2), diffuse reflecting type (ODM98) |
| BN-RM95-50C-PP | ± 0.5% (k=2), regular reflecting type (plan, very low scattering mirror) |
Ordering Information
| Model | Item No.: | Description |
| ISMP-50-R-V01 | 102731 | Light reflectance measurement instrument with sphere and electronic device, connection cables, manual (English) |
| BN-RM95-50C-PP | 102704 | Diffuse reflectance standard with calibration certificate |
| BN-R98-D2C-PP | 102697-1 | Mirror reflectance standard with calibration certificate |